| Statement of Deficiencies | (X1) Provider/Supplier/CLIA Identification Number 01D0300027 | (X3) Date Survey Completed 06/17/2025 |
| Name of Provider or Supplier Bibb Medical Center | Street Address, City, State 208 Pierson Avenue, Centreville, AL | |
| For information on the provider's plan to correct this deficiency, please contact the provider or the state survey agency. | ||
| (X4) ID Prefix Tag | Summary Statement of Deficiencies |
| D6089 | LABORATORY DIRECTOR RESPONSIBILITIES CFR(s): 493.1445(e)(4)(i) (e)(4)(i) The proficiency testing samples are tested as required under subpart H of this part; This STANDARD is not met as evidenced by: Based on proficiency testing (PT) desk reviews of the CASPER Reports 0153D and 0155D (Individual Laboratory Profiles from the Centers of Medicare and Medicaid Services [CMS]), and PT evaluation reports from the American Proficiency Institute (API), the laboratory director failed to ensure the laboratory had successful participation in an HHS approved proficiency testing program for Unexpected Antibody Detection in two out of three 2024 - 2025 API PT events. Refer to D2172 |