Statement of Deficiencies (X1) Provider/Supplier/CLIA Identification Number 01D0300027 (X3) Date Survey Completed 07/28/2021
Name of Provider or Supplier Bibb Medical Center Street Address, City, State 208 Pierson Avenue, Centreville, AL
For information on the provider's plan to correct this deficiency, please contact the provider or the state survey agency.
(X4) ID Prefix Tag Summary Statement of Deficiencies
D5439 CALIBRATION AND CALIBRATION VERIFICATION
CFR(s): 493.1255(b)

Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following: Perform and document calibration verification procedure - (b)(1) Following the manufacturer's calibration verification instructions; (b)(2) Using the criteria verified or established by the laboratory under ยง493.1253(b)(3)-- (b)(2)(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and (b)(2)(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and (b)(3) At least once every 6 months and whenever any of the following occur: (b)(3)(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes. (b)(3)(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance. (b)(3)(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem. (b)(3)(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification.


This STANDARD is not met as evidenced by:
Based on a review of the iStat calibration verification records and an interview with Testing Personnel #7, the laboratory failed to perform calibrations verifications at least every six months as per CLIA regulations. This was noted on one out of four calibration verifications reviewed for 2019 - 2021. The findings include: 1. A review of the iStat records revealed no calibration verification in December of 2019. The previous calibration verification was performed on 07/12/2019 and the next calibration verification was performed on 06/28/2020. 2. During an interview at 1:45 PM on 07/28/2021, Testing Personnel #7 confirmed the laboratory did not performed calibration verification in December 2019.