| Statement of Deficiencies | (X1) Provider/Supplier/CLIA Identification Number 01C0001008 | (X3) Date Survey Completed 10/11/2023 |
| Name of Provider or Supplier Birmingham Outpatient Surgery Center, Ltd | Street Address, City, State 2720 University Boulevard, Birmingham, AL | |
| For information on the provider's plan to correct this deficiency, please contact the provider or the state survey agency. | ||
| (X4) ID Prefix Tag | Summary Statement of Deficiencies
(Each deficiency should be preceded by full regulatory or LSC identifying information) |
| K0325 | Alcohol Based Hand Rub Dispenser (ABHR) CFR(s): NFPA 101 Alcohol Based Hand Rub Dispenser (ABHR) ABHRs are protected in accordance with 8.7.3.1, unless all conditions are met: o Corridor is at least 6 feet wide. o Maximum individual dispenser capacity is 0.32 gallons (0.53 gallons in suites) of fluid and 18 ounces of Level 1 aerosols. o Dispensers shall have a minimum of 4-foot horizontal spacing. o Not more than an aggregate of 10 gallons of fluid or 1135 ounces of aerosol are used in a single smoke compartment outside a storage cabinet, excluding one individual dispenser per room. o Storage in a single smoke compartment greater than 5 gallons complies with NFPA 30. o Dispensers are not installed within 1 inch of an ignition source. o If floor is carpeted, the building is fully sprinkler protected. o ABHR does not exceed 95% alcohol. o Operation of the dispenser shall comply with Section 20.3.2.6(11) or 21.3.2.6(11). o ABHR is protected against inappropriate access. 21.3.2.6, 8.7.3.1, CFR 416.44 This STANDARD is not met as evidenced by: Based on observation, the facility failed to install an Alcohol-Based Hand-Rub (ABHR) Dispenser per the requirements of: 2012 NFPA 101, 21.3.2.6 (8) This deficiency affects 1 of 6 smoke compartments. Findings include: During a tour of the facility, the surveyor observed an ABHR Dispenser was mounted 2’-0" directly above a sconce night light (ignition source) on the second floor in "Patient Room 1/3". A member of maintenance staff was present when this deficiency was identified. . |